Preparing for a measurement
The measurement must be performed on a patient who is lying down in order ensure
equal blood pressure among all limbs.
The patient should lie down for approx. 5 minutes before taking a measurement.
Fixing the cuff
The cuffs are colour coded. Affix the cuffs on both upper arms and both ankles
according to their colour coding (matching the symbols next to the air connection
sockets on the device).
The upper arm cuffs should be affixed in such a way that the lower edge of the cuff
is approx. 2-3 cm above the elbow joint. The marking on the cuff should be on the
arteria brachialis.
The metal ring should never be affixed over the arteries, as this may lead to incorrect
measurement values.
The ankle cuffs should be affixed in such a way that the lower edge of the cuff is
approx. 1-2 cm above the ankle. The marking on the cuff should be on the arteria
tibialis. Make sure that the cuff is tightly affixed to the ankle.
The device should only be used with original boso cuffs CA04 or CL04.
Type Circumference Order no.
CA04 22 - 42 cm 143 - 4 - 768
CL04 18 - 38 cm 143 - 4 - 769